eChip® – The Oximetry Module
- Accuracy proven by Bickler Laboratory of University of California San Francisco.
- Small Size allows it to fit into small handheld devices.
- Low Power Consumption which facilitating long battery life in handheld devices.
- Signal Index (Perfusion level) is providing a proportional to the perfusion level of the patient
- Plethysmogram is automatically scaled and offset in real time to fit into graphical display. A real-time pulse bar output is provided to indicate the strength of pulse.
- 3 response modes selection; Sensitive. Stardard & Stable.
- UART Communication which provided a flexible and straightforward data output making it implement in the host system easily
- A real-time warning and error reporting system, allowing the host to monitor whether eChip® is working properly. Messages and warnings are provided for abnormal operation.
Description
- Accuracy proven by Bickler Laboratory of University of California San Francisco.
- Small Size allows it to fit into small handheld devices.
- Low Power Consumption which facilitating long battery life in handheld devices.
- Signal Index (Perfusion level) is providing a proportional to the perfusion level of the patient
- Plethysmogram is automatically scaled and offset in real time to fit into graphical display. A real-time pulse bar output is provided to indicate the strength of pulse.
- 3 response modes selection; Sensitive. Stardard & Stable.
- UART Communication which provided a flexible and straightforward data output making it implement in the host system easily
- A real-time warning and error reporting system, allowing the host to monitor whether eChip® is working properly. Messages and warnings are provided for abnormal operation.
| Parameter | Range | Unit | ||
| Min. | Typ. | Max. | ||
| SpO2 | 0 | – | 100 | % |
| SpO2 accuracy70 – 99%
<70% |
–
– |
±2
Monotonic |
–
– |
% |
| SpO2 resolution | – | 1 | – | % |
| Pulse rate range | 30 | – | 249 | BPM |
| Pulse rate resolution | ±3 | BPM | ||
| Perfusion range | 0.05 | – | 20 | % |
| SpO2 averagingStable
Normal Sensitive |
–
– – |
16
12 6 |
–
– – |
Pulse |
| Motion artifact | – | Not Specified | – | / |
| Signal index | – | NormalReduced
Poor |
– | |
| Pulse wave | 0 | – | 255 | LSB |
| Pulse bar | 0 | – | 15 | LSB |
| Other specifications | ||||
| Parameter | Range | Unit | ||
| Min. | Typ. | Max. | ||
| Operating Temperature: | -20 | – | +60 | °C |
| Storage Temperature | -30 | – | +70 | °C |
| Operating humidity (Non Condensing) | 0 | – | 90 | % |
| Storage humidity (Non Condensing) | 0 | – | 90 | % |
| Dimensions | – | 18 x 31 x 6 | – | mm |
| Weight | – | 2 | – | g |